Optimizing gradient conditions in online comprehensive two-dimensional reversed-phase liquid chromatography by use of the linear solvent strength model

Authors
Publication date 09-2017
Journal Journal of Separation Science
Volume | Issue number 40 | 18
Pages (from-to) 3612-3620
Number of pages 9
Organisations
  • Faculty of Science (FNWI) - Van 't Hoff Institute for Molecular Sciences (HIMS)
Abstract
The linear solvent strength model was used to predict coverage in online comprehensive two-dimensional reversed-phase liquid chromatography. The prediction model uses a parallelogram to describe the separation space covered with peaks in a system with limited orthogonality. The corners of the parallelogram are assumed to behave like chromatographic peaks and the position of these pseudo-compounds was predicted. A mix of 25 polycyclic aromatic compounds were used as a test. The precision of the prediction, span 0-25, was tested by varying input parameters, and was found to be acceptable with root mean square errors of 3. The accuracy of the prediction was assessed by comparing with the experimental coverages. Less than half of experimental coverages were outside prediction +/- 1 x root mean square error and none outside prediction +/- 2 x root mean square error. Accuracy was lower when retention factors were low, or when gradient conditions affected parameters not included in the model, e.g. second dimension gradient time affects the second dimension equilibration time. The concept shows promise as a tool for gradient optimization in online comprehensive two-dimensional liquid chromatography, as it mitigates the tedious registration and modeling of all sample constituents, a circumstance that is particularly appealing when dealing with complex samples.
Document type Article
Note With supplementary file
Language English
Published at https://doi.org/10.1002/jssc.201700239
Permalink to this page
Back