A comparative study of the Osstell™ versus the Osstell Mentor™ to evaluate implant stability in human cadaver mandibles

Authors
Publication date 2013
Journal Journal of Oral Rehabilitation
Volume | Issue number 40 | 10
Pages (from-to) 774-779
Organisations
  • Faculty of Dentistry (ACTA)
  • Faculty of Medicine (AMC-UvA)
Abstract
The aim of this study was to assess the primary stability of dental implants by resonance frequency analysis (RFA) using the Osstell™ and Osstell Mentor™ devices and to investigate the reproducibility and comparability of the results obtained with these devices. Twenty-four Straumann implants (Straumann AG, Basel, Switzerland) were placed in the anterior mandible of 12 fresh edentulous human cadaver mandibles. The implant stability quotients (ISQs) were measured with the Osstell™ and Osstell Mentor™ when implants were inserted at 50% of their length and following their complete insertion. The Osstell™ device measured lower scores compared with the Osstell Mentor™. This was significant for the full position (mean difference = 9·9), t (11) = 7·4, P < 0·001 and for the halfway position (mean difference = 5·9), t (11) = 2·41, P = 0·03. In con-clusion, the Osstell™ produced relatively lower ISQ scores than the Osstell Mentor™.
Document type Article
Language English
Published at https://doi.org/10.1111/joor.12093