A comparative study of the Osstell™ versus the Osstell Mentor™ to evaluate implant stability in human cadaver mandibles
| Authors |
|
|---|---|
| Publication date | 2013 |
| Journal | Journal of Oral Rehabilitation |
| Volume | Issue number | 40 | 10 |
| Pages (from-to) | 774-779 |
| Organisations |
|
| Abstract |
The aim of this study was to assess the primary stability of dental implants by resonance frequency analysis (RFA) using the Osstell™ and Osstell Mentor™ devices and to investigate the reproducibility and comparability of the results obtained with these devices. Twenty-four Straumann implants (Straumann AG, Basel, Switzerland) were placed in the anterior mandible of 12 fresh edentulous human cadaver mandibles. The implant stability quotients (ISQs) were measured with the Osstell™ and Osstell Mentor™ when implants were inserted at 50% of their length and following their complete insertion. The Osstell™ device measured lower scores compared with the Osstell Mentor™. This was significant for the full position (mean difference = 9·9), t (11) = 7·4, P < 0·001 and for the halfway position (mean difference = 5·9), t (11) = 2·41, P = 0·03. In con-clusion, the Osstell™ produced relatively lower ISQ scores than the Osstell Mentor™.
|