Biomechanical evaluation of fixation of comminuted olecranon fractures: one-third tubular versus locking compression plating

Open Access
Authors
  • P. Kloen
Publication date 04-2010
Journal Archives of orthopaedic and trauma surgery
Volume | Issue number 130 | 4
Pages (from-to) 459-464
Organisations
  • Faculty of Medicine (AMC-UvA)
Abstract
New concepts in plate fixation have led to an evolution in plate design for olecranon fractures. The purpose of this study was to compare the stiffness and strength of locking compression plate (LCP) fixation to one-third tubular plate fixation in a cadaveric comminuted olecranon fracture model with a standardised osteotomy. Five matched pairs of cadaveric elbows were randomly assigned for fixation by either a contoured LCP combined with an intramedullary screw and unicortical locking screws or a one-third tubular plate combined with bicortical screws. Construct stiffness was measured by subjecting the specimens to cyclic loading while measuring gapping at the osteotomy site. Construct strength was measured by subjecting specimens to ramp load until failure. There was no significant difference in fixation stiffness and strength between the two fixation methods. All failures consisted of failure of the bone and not of the hardware. Contoured LCP and intramedullary screw fixation can be used as an alternative treatment method for comminuted olecranon fractures as its stiffness and strength were not significantly different from a conventional plating technique
Document type Article
Language English
Published at https://doi.org/10.1007/s00402-009-0980-z
Downloads
339541.pdf (Final published version)
Permalink to this page
Back