Time domain algorithm for accelerated determination of the first order moment of photo current fluctuations in high speed laser Doppler perfusion imaging

Open Access
Authors
  • M. Draijer
  • E. Hondebrink
  • T. van Leeuwen
  • W. Steenbergen
Publication date 2009
Journal Medical & Biological Engineering & Computing
Volume | Issue number 47 | 10
Pages (from-to) 1103-1109
Organisations
  • Faculty of Medicine (AMC-UvA)
Abstract
Advances in optical array sensor technology allow for the real time acquisition of dynamic laser speckle patterns generated by tissue perfusion, which, in principle, allows for real time laser Doppler perfusion imaging (LDPI). Exploitation of these developments is enhanced with the introduction of faster algorithms to transform photo currents into perfusion estimates using the first moment of the power spectrum. A time domain (TD) algorithm is presented for determining the first-order spectral moment. Experiments are performed to compare this algorithm with the widely used Fast Fourier Transform (FFT). This study shows that the TD-algorithm is twice as fast as the FFT-algorithm without loss of accuracy. Compared to FFT, the TD-algorithm is efficient in terms of processor time, memory usage and data transport
Document type Article
Language English
Published at https://doi.org/10.1007/s11517-009-0537-x
Downloads
320501.pdf (Final published version)
Permalink to this page