A thin tracheal silicone washer to solve periprosthetic leakage in laryngectomies: direct results and long-term clinical effects
| Authors |
|
|---|---|
| Publication date | 2008 |
| Journal | The Laryngoscope |
| Volume | Issue number | 118 | 4 |
| Pages (from-to) | 640-645 |
| Organisations |
|
| Abstract |
Objectives: Assessment of the immediate results and long-term clinical effects of a thin silicone washer placed behind the tracheal flange of voice prostheses to treat periprosthetic leakage.
Patients and Methods: Three year retrospective analysis of 32 laryngectomized patients with 107 periprosthetic leakage events (PLEs). Custom-made silicone washers (outer diameter 18 mm, inner diameter 7.5 mm, thickness 0.5 mm) were placed behind the tracheal flange either in combination with prosthesis replacement or later. Results: There was immediate resolution of periprosthetic leakage in 88 PLEs (median, 38 d; mean, 53 d; range, 8-330 d) and in 6 PLEs with the washer still in situ at the date of analysis (median, 75; mean, 97 d; range, 38-240 d). There was no resolution for periprosthetic leakage in 13 PLEs. Thus, in total, 94 of 107 PLEs (88%) were successfully resolved. In 29 of 32 (91%) patients, the washer resolved the problem at least in one PLE successfully. Twelve of 32 patients, including all 3 with washer failures, also required other interventions to ultimately solve the problem. The vast majority of patients (80%) did not consider placement of the washer to be inconvenient. Conclusions: In consideration of the high success rate and limited inconvenience for patients, this simple thin silicon washer application provides a good first option for the treatment of periprosthetic leakage. |
| Document type | Article |
| Language | English |
| Published at | https://doi.org/10.1097/MLG.0b013e31816067d5 |
| Downloads |
authors final version
(Accepted author manuscript)
|
| Permalink to this page | |