Postcolumn derivatization of amino acids using reaction flow chromatography columns with fluorescence detection: A fast new approach to selective derivatization techniques

Authors
  • S. Pravadali-Cekic
  • A. Jones
  • A.A. Kazarian
  • B. Paull
  • A. Soliven
  • H. Ritchie
  • M. Camenzuli
  • L. Leung
  • G.R. Dennis
  • R.A. Shalliker
Publication date 2017
Journal JOURNAL OF LIQUID CHROMATOGRAPHY RELATED TECHNOLOGIES
Volume | Issue number 40 | 2
Pages (from-to) 83-92
Organisations
  • Faculty of Science (FNWI) - Van 't Hoff Institute for Molecular Sciences (HIMS)
Abstract
Reaction flow (RF) chromatography with fluorescamine reagent and fluorescence detection (FLD) was used for the analysis of amino acids. The performance of RF chromatography was tested against several optimized conventional postcolumn derivatization (PCD) methods. RF columns achieved greater sensitivity compared to conventional PCD methods, without the need for reaction loops, which resulted in more efficient separations. The RF-PCD method also achieved limits of detection (LOD) from the low picomole to subnanomole range. The calibration data of the RF-PCD technique yielded R2 ≥ 0.99 and % relative standard deviation in peak areas ranging from 0.34% to 5%. Through reaction flow chromatography, multiplexed detection was also achieved allowing the monitoring and analysis of derivatized and nonderivatized flow streams simultaneously.
Document type Article
Note With supplementary file
Language English
Published at https://doi.org/10.1080/10826076.2017.1285312
Permalink to this page
Back